Super Systems 1500 VRSD User Manual

Page 10

Advertising
background image

Model 1500 VRSD Operations Manual

Super Systems Inc.

Page 10 of 28

13.0

Verify the 11.3% (Zero) calibration

13.1

Leave the sensor in the 11.3%RH calibration salt.

13.2

Record the temperature and the dew point of the sensor. Since there is

no display on the instrument, you will need to measure the voltage from the

microprocessor board and translate that into the appropriate measurement.

13.2.1

To determine the sensor temperature, measure the DC voltage

between pins 5(+) and 9(-) on the microprocessor board. The

voltage will be between 0 and 1VDC. Use the chart in Appendix C to

determine the temperature in °F, or use Appendix D to determine

the temperature in °C.

13.2.2

Temporarily record the sensor temperature.

13.3

Look up this temperature in Appendix “A” (Determining the Dew Point

in °F) or Appendix “B” (Determining the Dew Point in °C). Appendix A will

show the temperature values in Fahrenheit, and Appendix B will show the

temperature values in Celsius.

13.4

Next to the appropriate sensor temperature, note the number in the

corresponding column titled “11.3%”. This will match the measured dew

point, which can be verified wherever the dew point is displayed.

14.0

Determine the acceptability of the reading.

14.1

The value printed on the chart in Appendix A is a theoretical value, and

some variation can be expected. When a calibration is performed at SSi, we

certify (in writing) that the unit displays within +/- 1 degree of the theoretical

value after it has been calibrated. We would not consider a calibration to be

successful unless it is within +/- 1 degree, however in the case of a field

calibration, this degree of accuracy may or may not be required. The degree

of accuracy that is acceptable is determined by the policy of the person

performing the calibration.

NOTE: Keep in mind that the VRSD only displays whole numbers, and not

tenths of a degree. Therefore, a reading of 18°F could be as low as 17.50 or

as high as 18.49.

15.0

Allow the sensor to achieve equilibrium at ambient atmosphere.

15.1

After the 11.3% (Zero) calibration has been completed, remove the sensor

from the calibration salt and replace the cap.