Super Systems 1500 VRSD User Manual

Page 9

Advertising
background image

Model 1500 VRSD Operations Manual

Super Systems Inc.

Page 9 of 28

value after it has been calibrated. We would not consider a calibration to be

successful unless it is within +/- 1 degree; however, in the case of a field

calibration, this degree of accuracy may or may not be required. The degree

of accuracy that is acceptable is determined by the policy of the person

performing the calibration.

NOTE: Keep in mind that the VRSD only displays whole numbers, and not

tenths of a degree. Therefore, a reading of 65°F could be as low as 64.50 or

as high as 65.49.

9.0

Allow the sensor to achieve equilibrium at ambient atmosphere.

9.1

After the 75.3% (Span) calibration has been completed, remove the sensor

from the calibration salt and replace the cap on the salt.

9.2

Leave the sensor probe in the gland and while the unit is still on, allow it to

achieve equilibrium at the ambient atmosphere in the room. This is

accomplished by simply leaving the sensor exposed to ambient air for

between two and three minutes. You will know when this has been

accomplished when the numbers on the display begin to stabilize.

10.0

Install the sensor probe into the 11.3% salt solution.

10.1

Remove the cap of the 11.3% salt solution and install the sensor gland

(with the sensor) into the salt solution. To increase the life of the calibration

salts, an effort should be made to minimize the amount of time that the salt

solution is exposed to the ambient air.

10.2

Turn the unit off.

11.0

Allow the sensor to reach equilibrium with the calibration salt.

11.1

With the power to the unit still turned off, leave the sensor in the

calibration salt for a minimum of twenty-four (24) hours. It is acceptable to

leave the sensor in the salt solution for a longer period of time, even a few

days, if desired.

12.0

Begin the 11.3% (Zero) calibration process

12.1

After leaving the sensor in the salt for at least twenty-four (24) hours, turn

the unit on. The reading on the display is not important at this point.

12.2

Simultaneously press the “11.3%” and “Calibration” buttons on the

microprocessor board.

Advertising