Model 1500 VRSD Operations Manual

Super Systems Inc.

Page 5 of 28

Warnings

Although it is intended for use in an industrial environment, this is a sensitive piece of

analysis equipment. Care should be taken not to operate it in a manner inconsistent

with its intended use.

Moisture (water) cannot be allowed to enter the unit. If water is present in the

sample gas, use an in-line dryer for sample conditioning. In the event that the

sensor becomes wet, use an inert gas (Nitrogen or Argon) to dry the inside of the

instrument. Under no circumstances should Methane (Natural Gas) be used to dry

the unit.

The analyzer must be stored at ambient temperature (65-80°F) for at least four

hours prior to operation.

An in-line dryer for sample conditioning should be used for exothermic and

combustion applications.

This unit is not designed to measure the dew points in corrosive gasses, such as

Ammonia, S0

3

, Chlorine, and HCL.


IMPORTANT!

Failure to comply with these conditions may cause damage to the unit that will not be

covered under the warranty. Super Systems, Inc. is not responsible for damage to this

unit caused by disregard of these warnings, neglect, or misuse.

Specifications

Measurement Range:

0 to +80°F (-17.8 to +27°C)

Temperature Range:

0 to +120°F (-18 to +49°C)

Accuracy:

+/- 1°F (+/- 0.5°C)

Zero and Span Drift:

+/- 1°F (+/- 0.5°C)

Power Supply:

115 VAC 60Hz (Optional 220VDC)

Display Type:

3.5” Color Touch Screen

Display Resolution:

+/- 1°F (+/- 0.1°C)

Digital Communications:

(2) RS485, RS232, Ethernet

Control / Retransmission Output:

4 – 20 mA (2 Outputs)

Alarms:

Up to 3 Alarm Relays

Dimensions:

20” x 16” x 10” (508mm x 406mm x 254mm)

Weight:

34.4 lbs. (15.6 kg)