Operation, Returning the unit to ssi, Operation returning the unit to ssi – Super Systems 1500 VRSD User Manual

Page 12

Advertising
background image

Model 1500 VRSD Operations Manual

Super Systems Inc.

Page 12 of 28

Operation

To obtain consistent accurate readings from the unit, be sure that the element in the

bowl filter on the side of the instrument is clean and functional. Not only will this

ensure that the sample reading is not abnormally high (since soot tends to trap

moisture), but it will also prevent soot and other contaminants from entering the unit

and damaging the sensor. The optimum flow rate of the sample gas should be between

1.5 and 2.0 Standard Cubic Feet per Hour (SCFH), although a flow rate as low as 1.0

SCFH is acceptable. If the unit is reading less than 1.0 SCFH, verify that there are no

obstructions to the flow such as a clogged sample line or a poorly adjusted knob on the

1500 VRSD flow meter.

Heat Treat Furnace Sampling:

A gas sample may be extracted from a process using the

built-in pump. The sample tube from which the sample is taken out of the furnace

should extend into the furnace past the HOT face of the refractory. For accurate

results, a designated sample port should be used to extract the sample. SSi offers a

sample port assembly (part number 20263) which is ideal for this purpose.

Endothermic Generator Sampling:

For applications under pressure, the pump should be

switched off and the flow controlled by the small restriction valve on the flow meter. A

flow rate between 1.5 and 2.0 SCFH is ideal. The sample should be taken from the

endothermic gas manifold after the gas has been cooled. NOTE: Allow the sample port

“to blow out any soot” before connecting the sample tube. Failure to do so will

unnecessarily coat the sample tubing assembly and possibly some internal components

with soot, resulting in inaccurate readings and exposing the sensor to potential

damage.

Returning the Unit to SSi

This analyzer contains some components that may require periodic replacement based

on the amount of use that the unit experiences and the methods in which it is used. If

service on the unit is necessary, it should be sent back to Super Systems, Inc. for repair.

To minimize damage to the mounting feet on the enclosure, it is possible to unscrew

them and rotate them 180 degrees (so they point into the enclosure instead of away

from it). This will reduce the likelihood that they will be damaged during shipment. If

the original packaging is not available, the analyzer should be surrounded by impact-

absorbing materials and placed in a box. It is the responsibility of the shipper to ensure

that the unit arrives at SSi undamaged.

Advertising