If “Unidirectional” is chosen, the output will give pulses proportional to flow rate when the
flow rate is positive, and will not give them when the flow rate is negative.

If “Bidirectional” is chosen, the output will give pulses proportional to flow rate in both flow
directions.

7.5.2.2 Units

First volume units for which we want to get a number of pulses are selected.

7.5.2.3 Pulses

Then the pulses per unit of volume that we will obtain at the output are selected .

7.5.2.4 Duty Cycle

With the duty cycle width or percentage of time the pulse will be active during each cycle
can be determined

Active pulse is when the NPN transistor at the output (see 2.2.5) is ON.

7.5.3 Analog

output

The flow rate corresponding to 4 mA and 20 mA can be programmed and the calibration

for these two values can be made as well.

CONVERTER PROGRAMMING