1 sensor position, 2 straight pipe sections – Tecfluid MX4 For FLOMID Series User Manual

Page 8

Advertising
background image

8

4.2

Straight pipe sections

They are necessary before and after the sensor. The minimum distances are the
following:

Upstream

5

DN

Downstream

3

DN

In installations with turbulent flow it may be necessary to increase these distances.

4.1 Sensor

position

The most adequate position is with the electrodes in a horizontal plane. In this way,
deposits of particles on the electrodes are avoided.

INSTALLATION

4

Advertising