Step 3: high magnification microscope inspection, Pyramid probes: new core inspection – Cascade Microtech Pyramid Probes User Manual

Page 3

Advertising
background image

PN 153-690-3

www.cascademicrotech.com

Pyramid Probes: New Core Inspection • 3

Step 3: High Magnification Microscope Inspection

1. Place the core on a high-magnification fixture with the probe tips facing up.

2. Inspect the core using the 5x objective.

Components

Is there any damage to installed components?

Small scratches from the assembly process are acceptable.
Damage to the body of a component is not acceptable on new cores.

Inspection Area

Criteria

Probe Face

Are there any scratches on the probe face?

New cores will have no scratches greater
than 1000 μm.

Damaged component, not

acceptable

Scratch >1000 μm,

not acceptable

Scratch >1000μm,

not acceptable

Pyramid Probes: New Core Inspection