Iii-b. operation checks, Iii-c. shutdown procedure, Iv-a. lubrication – Eagle Pump & Compressor A100 ANSI PUMPS User Manual

Page 6: Iv-b. stufng box, Iv-c. vibration, Iv-d. alignment—final, Iv-e. performance

Advertising
background image

6

b. A high quality turbine type oil with rust and oxidation

inhibitors should be used. Under normal operating

conditions, an oil of 300 SSU viscosity at 100° F.

(approximately SAE 20) should be used. Fill oiler bottle

and replace in oiler housing. Repeat until oil remains

visible in bottle. Do not add oil through the vent or

breather. Optional grease lubricated bearings are lubricated

at the factory and need lubrication only after 2,000

hours of operation, or every 3 months, which ever occurs

first. On pumps supplied with greased–for–life bearings,

no additional lubrication is required for the life of the

bearing.

Oil Mist lubrication of bearings is available on A100–X

frames. This can be supplied at time of order, or field

conversion can be done. Refer to the instructions of the

manufacturer of the Oil Mist System.

c. Priming—Pump and suction piping must be full of liquid

before pump is started. Usually suction supply will be

primed when shutoff valves are opened, if pump is below

suction supply. If suction supply is below pump, priming

by other means, such as a foot valve or ejector, will be

required.

d. Free Rotation—Rotate shaft by hand to make sure it is

free. Drag from packing or seal is normal but, if pump

cannot be rotated by hand or binding or rubbing is noticed,

correct before starting pump.

2. Startup

a. Valves—Be sure suction valve is fully open. Normally,

discharge valve should be at least partially closed for flow

control.

b. Rotation Check—if not already done, uncouple the unit

and jog the motor to check for proper rotation (refer to

Section II–E–2, page 5).

III-B. Operation Checks

Inspect pump carefully and frequently during the first few hours

of operation. If packing runs hot, shut pump down, allow box

to cool, loosen gland if necessary. (Do not loosen gland until

packing has cooled.) Mechanical seal may weep slightly, but

should “run–in” in a few hours. Be sure all auxiliary lines

(cooling, flushing, sealing, etc.) are functioning properly. Check

pump bearings for excessive heating. Bearing housing operating

temperatures vary depending on a wide variety of conditions.

However, normal temperatures should be 120° to 180° F. (49° to

82° C). A change of temperature can indicate a problem, and

operating temperatures outside of the normal range should be

addressed with manufacturer. Check complete unit for excessive

vibration and unusual noises. Do not run pump at greatly

reduced flow because damage can result.

III-C. Shutdown Procedure

Back flow through pump will cause reverse rotation. If backflow

is excessive, and there is a possibility of the pump being turned

on during this period of reverse rotation, then precautions

should be taken to prevent the backflow. This can be done

by installing a check valve in the discharge line, or by closing

a discharge valve immediately prior to shutting down the

pump. NOTE: IT IS NOT RECOMMENDED THAT THE

PUMP RUN LONGER THAN ABSOLUTELY NECESSARY

AGAINST A CLOSED DISCHARGE VALVE.

SECTION IV—PREVENTIVE AND CORRECTIVE MAINTENANCE

IV-A. Lubrication

Oil lubricated units require that oil be visible in the reservoir at

all times. Oil should be changed every 4000 hours of operation.

Grease lubricated units should be regressed every 2000 hours

or 3 month intervals, whichever occurs first. Use a sodium

or lithium grease and fill until grease comes out grease relief

fittings. Follow motor and coupling manufacturers’ lubrication

instructions.

IV-B. Stufng Box

1. Packing Stuffing Box:

Periodically inspect stuffing box to see that there is sufficient

leakage to lubricate the packing and maintain a cool box. Never

restrict the leakage from the packing as this will cause damage

to both packing and shaft sleeve. Draw up gland nuts slowly

and evenly and only while pump is running.

After pump has been in operation for some time and the packing

has been completely “run–in”, a leakage of 40 to 60 drops per

minute of the liquid should be allowed to flow from the stuffing

box at all times for cooling and lubricating the packing and

shaft sleeve.

2. Stuffing Boxes with Mechanical Seal:

Although this type of box requires no attention, a periodic

inspection of the circulating lines will ensure that they do not

become clogged.

IV-C. Vibration

It is good practice to periodically monitor vibration of the

pump. Normally, the vibration level will be well within accepted

standards. Of equal importance is that the vibration level not

increase. If a problem with vibration is encountered, refer to

Trouble Shooting, Section VII.

IV-D. Alignment—Final

Alignment should be checked after unit has reached operating

temperature, following startup. Repeat alignment procedures

outlined in Section II–E. Check alignment again after one week

of operation.

IV-E. Performance

If performance deteriorates, refer to Trouble Shooting, Section

VII.

Advertising