Gas selection – Ion Science GasCheck Tesla User Manual

Page 13

Advertising
background image

GASCHECK TESLA MANUAL

Ion Science Ltd

Page 13 of 25

Unrivalled Detection. www.ionscience.com

Functions

Gas selection
GasCheck Tesla is calibrated using Helium gas at manufacture, however other
gases can be detected and measured using a gas table within the instrument.
Although GasCheck Tesla can't differentiate between gases it can quantify other
gases using response factors stored in its internal gas table. Gas groups 0, 1 2, 3,
and 4 offer a generic sensitivity.

The following list indicates response factors for specific gases however for advice on
gases not listed please contact Ion Science Ltd for advice.

Name

Abbreviation

Minimum sensitivity

cc/sec (ml/sec)

Minimum sensitivity

cc/sec (ml/sec)

Hydrogen

H2

7.7 E-6 cc/sec

0.0000077 cc/sec

Helium

He

1.0 E-5 cc/sec

0.000010 cc/sec

Refrigerant R12

R12

2.7 E-5 cc/sec

0.000027 cc/sec

Refrigerant R1301

R1301

2.4 E-5 cc/sec

0.000024 cc/sec

Refrigerant R134a

R134a

5.8 E-5 cc/sec

0.000058 cc/sec

Refrigerant R22

R22

2.6 E-5 cc/sec

0.000026 cc/sec

Refrigerant R11

R11

3.2 E-5 cc/sec

0.000032 cc/sec

Sulphur Hexaflouride

SF6

2.2 E-5 cc/sec

0.000022 cc/sec

Carbon dioxide

CO2

4.0 E-5 cc/sec

0.000040 cc/sec

Methane

CH4

2.9 E-5 cc/sec

0.000029 cc/sec

Argon

Ar

3.5 E-5 cc/sec

0.000035 cc/sec

Oxygen

O2

2.9 E-4 cc/sec

0.000290 cc/sec

Refrigerant R502

R502

3.0 E-5 cc/sec

0.000030 cc/sec

Refrigerant R404a

R404a

3.2 E-5 cc/sec

0.000032 cc/sec

Refrigerant R407c

R407c

3.3 E-5 cc/sec

0.000033 cc/sec

Refrigerant R410a

R410a

3.2 E-5 cc/sec

0.000032 cc/sec

Refrigerant R507

R507

3.8 E-5 cc/sec

0.000038 cc/sec


To select an alternative gas use the 'Up and Down' keys to select the desired gas
and then press the 'ENTER' key to confirm the selection, at which point a tick will
appear. Press the 'ENTER or Esc' key to exit into the 'Main menu'.

If the gas to be detected does not appear in the instruments internal gas table then
one of the five 'Gas Groups' can be selected. 'Gas Groups' have a variation of
sensitivities. (Group 1 being high sensitivity through to Group 5 being low sensitivity)
Contact Ion Science Ltd for advice. Gas groups allow the instrument to calculate
leak rates with reasonable accuracy.


Advertising
This manual is related to the following products: