Energizer ATX612 User Manual

Page 13

Advertising
background image

20

21

• Remove the battery pack from the instrument as

described in Section 8.2.

• Remove the two screws which hold the battery module

in place as shown in Figure 4.

• Lift the Nicad battery module from the battery pack.

• Place the new Nicad module in the battery pack.

• Replace the two screws which hold the Nicad module

in place.

8.2.2

REPLACING ALKALINE BATTERY CELLS

The alkaline battery pack holds 6 C-cell batteries. To
replace the alkaline battery cells:

• Remove the battery pack from the instrument as

described in Section 8.2.

• Remove the alkaline cells from the battery pack.

• Insert the new cells making sure to observe proper

polarity.

NOTE: Proper polarity of the alkaline battery cells is

identified by the molded “+” and “-” symbols in
the bottom of the battery pack. In addition the
positive (+) battery contact is identified by the red
polarizing tab.

8.3 O

PENING THE

I

NSTRUMENT FOR

S

ERVICE

To open the instrument for service:

• Remove the battery pack as described in Section 8.2.

• Remove the four screws from the bottom of the

instrument chassis.

• Gently lift the case top from the chassis.

8.4 I

NSTALLING OR

C

HANGING

S

ENSORS

To change sensors in the ATX612:

• Open the instrument as described in Section 8.3.

• Remove the three screws which hold the sensor manifold

in place.

• Gently lift the sensor manifold away from the chassis as

shown in Figure 5.

NOTE: It is not necessary to disconnect the sample tubing

from the manifold to replace the sensors.

8.1 C

LEANING

Wipe the outside of the instrument with a soft, clean cloth.
Never use solvents or cleaning solutions of any type.

8.2 C

HANGING THE

B

ATTERY

P

ACK

To change the ATX612 battery pack:

• Hold the instrument with battery pack facing up.

• Using the maintenance tool provided with the

instrument, turn counter-clockwise and loosen the two
hex screws in the battery pack.

• Remove the battery pack. (See Section 8.2.1 for

instructions of replacing the Nicad battery module)

• Place the battery pack on the instrument.

• Turn the hex screws clockwise until the screws are tight

and reach the stops. DO NOT OVER TIGHTEN.

8.2.1

REPLACING THE NICAD BATTERY MODULE

The Nicad cells within the battery pack may be replaced
when necessary by installing a new Nicad battery module.
To replace the Nicad battery module:

Figure 5

Figure 4

Battery Pack
Hex Screws

Nicad Battery
Module

Battery Pack

8.

M

AINTENANCE

Advertising