Atx612 o, Efault, Larm – Energizer ATX612 User Manual

Page 16: Ettings, Pecifications, Ptions, Rdering, Nformation

Advertising
background image

27

26

11.

D

EFAULT

A

LARM

S

ETTINGS

GAS

LOW ALARM

HIGH ALARM

O

2

19.5%

23.5%

LEL

10%

20%

CH

4

1.0%

1.5%

CO

35 PPM

70 PPM

H

2

S

10 PPM

20 PPM

SO

2

2.0 PM

4.0 PPM

NO

2

3.0 PPM

6.0 PPM

Cl

2

0.5 PPM

1.0 PPM

ClO

2

0.3 PPM

1.0 PPM

NOTE: Factory alarm settings may not coincide with

local regulations. Consult all appropriate local
regulations for appropriate alarm settings in
your region.

10.

S

PECIFICATIONS

C

ASE

:

Type 304 Stainless Steel

D

IMENSIONS

:

8.2”L x 3.7”W x 3.2”H
(208 X 94 X 81 mm)

W

EIGHT

:

3.4 lbs (1.5 kg)

S

ENSORS

:

Combustible Gases and Methane-
Catalytic
Oxygen and Toxic Gases-
Electrochemical

M

EASURING

R

ANGE

:

LEL (Combustible Gases)

0 to 100% LEL in 1% increments

CH

4

(Methane)

0 to 5% of volume in 0.1% increments

O

2

(Oxygen)

0 to 30% of volume in 0.1% increments

CO (Carbon Monoxide)

0 to 999 PPM (parts per million) in 1 PPM increments

H

2

S (Hydrogen Sulfide)

0 to 999 PPM (parts per million) in 1 PPM increments

SO

2

(Sulfur Dioxide)

0 to 99.9 PPM (parts per million) in 0.1 PPM increments

NO

2

(Nitrogen Dioxide)

0 to 99.9 PPM (parts per million) in 0.1 PPM increments

C1

2

(Chlorine)

0 to 99.9 PPM (parts per million) in 0.1 PPM increments

C1O

2

(Chlorine Dioxide)

0 to 99.9 PPM (parts per million) in 0.1 PPM increments

P

OWER

S

OURCE

: Rechargeable, replaceable nickel-cadmium battery

pack, or replaceable cell alkaline battery pack

B

ATTERY

L

IFE

:

With Nicad Battery Pack - 16 hours typical
With Alkaline Battery Pack - 20 hours typical

R

EADOUT

:

Alpha-Numeric Liquid Crystal Display

T

EMPERATURE

R

ANGE

:

-20ºC to 50ºC (-4ºF to 122ºF)

With H

2

S Sensor

-40ºC to 50ºC (-40ºF to 122ºF)

H

UMIDITY

R

ANGE

:

0% to 99% RH (Non-condensing)

S

TORAGE

T

EMPERATURE

: 0ºC to 20ºC (32ºF to 68ºF)

P

ART

N

UMBER

D

ESCRIPTION

(Q

TY

)

1705-0788-PPM

Combustible Sensor

1705-0129

Oxygen Sensor

1704-1898

Hydrogen Sulfide Sensor

1704-1880

Carbon Monoxide Sensor

1704-1914

Chlorine sensor

1704-4204

CLO

2

sensor

1704-1922

Nitrogen Dioxide Sensor

1704-1906

Sulfur Dioxide Sensor

1810-1386

Stainless Steel Extendible Probe - 6ft.

1810-1428

Polycarbonate Probe

1705-9494

Rechargeable Nicad Battery Pack

1705-9312

Replaceable Alkaline Battery Pack (6 C-cell)

1810-3259

Leather Carrying Case for ATX612

1810-0628

Shoulder Strap

1810-2921

Carrying Handle for ATX612

1810-1154

External Audible/Visual Alarm

1810-2146

External Vibrating Alarm

1810-2187

Cylinder Cal. Gas, H

2

S, CO, Pentane and Oxygen

1810-1576

Cylinder, Cal. Gas, Carbon Monoxide, Pentane and Oxygen

1810-2165

Cylinder, Cal. Gas, Carbon Monoxide, Methane and Oxygen

1810-1584

Cylinder, Cal. Gas, Zero Air

1810-2222

Cylinder, Cal. Gas, 5 PPM Sulfur Dioxide

1810-1758

Cylinder, Cal. Gas, 10 PPM Chlorine

1810-2219

Cylinder, Cal. Gas, 5 PPM Nitrogen Dioxide

1810-2509

Demand Flow Regulator with Pressure Gauge

12.

ATX612 O

PTIONS

& O

RDERING

I

NFORMATION

Advertising