Energizer ATX612 User Manual

Page 6

Advertising
background image

INDUSTRIAL
SCIENTIFIC

ATX612

OXYGEN

LEL

E

BATT

PPM

%

ON/OFF MODE

6

7

Hidden (+) Key

Alarm Speaker

On/Off/Mode Key

Ultra-bright Visual
Alarm Light Bar

Battery Status
Indicator

Hidden (-) Key

Display

Enter/Backlight Key

Charging Socket
(Nicad Versions)

Infrared Data Interface

Gas Sample Inlet
Fitting

Pump Exhaust Port

Shoulder Strap/
Optional Handle
Attachment

Battery Pack
Hex Screws

External Alarm Jack

Interchangeable Nicad
or Alkaline Battery
Pack

A dead battery will be fully recharged in 4.5 hours. The

Nicad battery pack has a LED charge status indicator to
show the current status when connected to the charger.
The LED indicator has five states as follows:

• Solid Amber:

Fast charge in progress

• Solid Green:

Fully charged, in trickle
charge mode

• Flashing Green:

Battery too hot/cold to be
fast charged, in trickle
charge mode

• Flashing Amber:

Instrument running, fast
charge mode

• Alternating Green/Amber: Trouble, electrical problem

in battery pack

A fully charged Nicad battery pack will typically operate a
four-gas configured ATX612 for up to 16 hours. When the
instrument is in the normal viewing mode, an eight-
segment battery status indicator continuously displays the
battery condition. Each segment represents approximately
2 hours of operating time. When the battery is fully
discharged, the display will read BATTERY FAIL and the
instrument will emit a short beep once a second (See
Section 4.6.6). Turn off the instrument and recharge the
battery pack when BATTERY FAIL appears.

WARNING: Recharge Nicad battery pack only in a

non-hazardous location.