Sensor view command (sv), Sensor select command (ss), Sensor delete command (sd) – Epson 9840 User Manual

Page 35

Advertising
background image

I N T E R F AC E M O D E L 9 8 4 0 V E R 6 . 0 . 2

Example:
@123SA
Acknowledge:
@123 This is the list of cell calibration data:
Ch A = S/N 123456, 100.00 Lb , 3.00150 mV/v,
10.00 V , Cal on Oct27-99, 49.532 Lb Shunt

Sensor View Command (SV)

The Sensor View (SV) command has the format SV. It returns information about the
currently selected cell including the serial number, the maximum load rating, the
millivolt per volt calibration constant, the excitation voltage used, the latest date of
calibration, and the shunt value recorded during that calibration. All other cells
currently in the unit are listed as well.

Example:
@123SV
Acknowledge:
@123 This is the list of cell calibration data:
Ch A = S/N 123456, 100.00 Lb , 3.00150 mV/v,
10.00 V , Cal on Oct27-99, 49.532 Lb Shunt
unused S/N 13368, 10.00 Lb , 3.00121 mV/v,
10.00 V, Cal on Oct27-99, 4.9292 Lb Shunt
unused S/N 63220, 100.00 Lb , 2.99984 mV/v,
10.00 V , Cal on Oct27-99, 49.381 Lb Shunt
Ch B = S/N 89991, 1000.0 Lb , 4.50015 mV/v,
5.00 V , Cal on Oct27-99, 486.45 Lb Shunt

Sensor Select Command (SS)

The Sensor Select (SS) command is used to select the calibration data for a stored cell
using the cell serial number. The format is SS(channel)(serial number)#.

Example:
@123SSA13368#
Acknowledge:
@123 This is the list of load cell calibration data:
unused S/N 123456, 100.00 Lb , 3.00150 mV/v, 10.00 V,
49.532 Lb Shunt
Ch A = S/N 13368, 10.00 Lb , 3.00121 mV/v, 10.00 V,
4.9292 Lb Shunt
unused S/N 63220, 100.00 Lb , 2.99984 mV/v, 10.00 V,
49.381 Lb Shunt
Ch B = S/N 89991, 1000.0 Lb , 4.50015 mV/v, 5.00 V,
486.45 Lb Shunt



Sensor Delete Command (SD)

The Sensor Delete (SD) command is used to remove the calibration data for a stored
cell from the list. The format is SD(serial number)#.

M O D E L 9 8 4 0 P G 3 4

P U B .

2 8 5 6 - 1 6

Advertising