I N T E R F AC E M O D E L 9 8 4 0 V E R 6 . 0 . 2

INTRODUCTION

The Model 9840 is a CE compliant and versatile precision instrument intended for the
digital readout of strain gage sensors such as load cells and torque cells. Optical
encoders are also supported.

Here is a quick listing of its features:

• Full

bi-polar

6 digit display (

±999,999) using two line, 20 character, vacuum

fluorescent display for clear, precise indication of measured quantities and limit status.


No knobs or dials. No internal jumpers. All adjustment and calibration is done using

user-friendly front panel menus. Keylock option to prevent accidental or unauthorized
changes to the unit’s setup.


• Measure

load and gross weight in pounds, kilograms, Newtons, PSI, MegaPascals,

kilo-pounds, kilo-Newtons, metric tons, mV/V, or grams. User defined base area for
PSI and MPa. Tare at any time. The sum of the loads on both channels is also
availabe.


• Measure

torque and gross torque in Pound-Inches, Newton-Meters, Ounce-Inches,

or mV/V. Tare at any time. The sum of the torques on both channels is also availabe.


• Capture the load peak and valley, display in pounds, kilograms, Newtons, PSI,

MegaPascals, kilo-pounds, kilo-Newtons, metric tons, mV/V, or grams. Reset at any
time.


• Capture the torque peak and valley, display in in Pound-Inches, Newton-Meters,

Ounce-Inches, or mV/V. Reset at any time.


• Measure

position and velocity with optional encoder input. Display in English units,

metric units, or in percent of a user defined base length. Reset at any time.


• Four contact closure Limit switches. You can set which signal drives the limit (load,

torque, peak, valley, position, or velocity), normally open / normally closed, set point,
and manual reset (latched) or automatic reset (not latched).


Print current readings of load, torque, peak, valley, position, and velocity using the

currently selected units at the touch of one button (optional).


Automatic identification of calibrated cells with auto-id option.

Calibration for cells using manufacturer’s milli-volt per volt calibration factor, 6 point

mass and milli-volt per volt, 2 or 5 point known mass, 6 point torque and milli-volt per
volt, 2 or 5 point known torque, or internal precision shunt calibration. A back panel
switch allows you to choose between two values used for the shunt calibration so
there is no need to open the unit.


Analog output of any signal (load, torque, peak, valley, position, velocity) with user

defined scale factor and offset. Large, buffered

± 10.00 volt output.

M O D E L 9 8 4 0 P G 4

P U B .

2 8 5 6 - 1 6