I N T E R F AC E M O D E L 9 8 4 0 V E R 6 . 0 . 2

Calibration by mV/V Mass 1 (CMVM1) command. The mass is entered using the
following format: CMVM1(mass value)#. Entering this command will result in the unit
prompting the user for the first milli-volt per volt command.

Example:
@123CMVM15.0005#
Acknowledge:
@123 Calibrate Mass 1 Command entered

Ready for mV/V Value CMVV1 or CE command

The CE or Calibration Escape command will, if entered, end this calibration without
calibrating the unit. The unit will remain calibrated according to the most recent
calibration or, if this is the first calibration of the unit, the unit will remain uncalibrated.

At this point the unit is prompting the user for the first milli-volt per volt value of the
calibration. The mV/V value is entered using the Calibration by mV/V Volt Command
with the following format: CMVV1(mV/V value)#.


Example:
@123CMVV11.9926#
Acknowledge:
@123 Calibrate mV/V 1 Command entered

Ready for Mass Value CMVM2 or CE command

The unit will continue to prompt the user for mass and milli-volt per volt values until the
six points have been entered. Then the user will see the following acknowledgement:

@123 Calibrate mV/V 6 Command entered

Ready for Mass Value CMVM0 or CE command

The user must now enter the CMVM0 command to finish the calibration. Note the load
cell should have no masses on it at this point since a shunt calibration is run.


Example:
@123CMVM0
Acknowledge:

@123 Calibrate Command - Reading for Shunt Check…

[10 seconds delay]

@123 Calibrate Command Completed

Ch A = S/N 123456, 1000.0 Lb , 4.50020 mV/v,

4.50020

mV/v,

4.50020

mV/V,

4.50020

mV/V

4.50020

mV/V

5.00 V , Cal on Oct27-99,

553.26 Lb Shunt

Calibrate by mV/V Torque and Volt Commands (CMVT and CMVV)

The Calibration by mV/V Torque (CMVT) and Volt (CMVV) works exactly like the
calibrate by mV/V Mass and Volt commands. These commands can have up to seven
parts each: CMVT1 through CMVT6 and CMVT0; and CMVV1 through CMVV6 and
CMVV0. You must have already started a calibration using the CB and CMV6
commands for any of these commands to function. To input the first torque issue the

M O D E L 9 8 4 0 P G 3 8