Sensor Installation: F-Series

Wiring

Figure 11

Micro Motion cable gland and heat shrink

4. For connection at the core processor housing, prepare shielded cable as follows (for armored cable,

omit steps d, e, f, and g):

a. Strip 4 1/2 inches (114 mm) of cable jacket.

b. Remove the clear wrap that is inside the cable jacket, and remove the filler material between the

wires.

c. Remove the foil shield that is around the insulated wires, leaving 3/4 inch (19 mm) of foil or braid

and drain wires exposed, and separate the wires.

d. Wrap the shield drain wire(s) around the exposed foil twice. Cut off the excess wire.

Figure 12

Wrapping the shield drain wires

e. Place the shielded heat shrink over the exposed shield drain wire(s). The tubing should completely

cover the drain wires.

f. Without burning the cable, apply heat (250 °F or 120 °C) to shrink the tubing.

4 1/2 in

(114 mm)

3/4 in

(19 mm)

7/8 in

(22 mm)

7/8 in

(22 mm)

Shielded heat shrink

Gland body

Gland nut

Gland clamping

insert