15

Installation

3.5.2 Installation of the cuvette/sample cell adapters

1. Open the cell compartment.

2. Select the correct adapter for the cuvette/sample cell type.

3. Insert the adapter so the arrow on top of the adapter points to

the left (

Figure 5 on page 16

) and the orientation tab fits the

groove in the compartment opening. The cuvette/sample cell
type imprint should be legible on the adapter (

Figure 4

).

Note: The arrow on top of the adapter indicates the direction of the
light beam path.

Figure 4 Cuvette/Sample cell adapters

1

10 mm square cuvette/sample cell adapter (A)

3

1 inch round cuvette/sample cell adapter (C)

2

1 inch Pour-Thru adapter (B)