4 enter variables, 5 save a free programming user program, Advanced operations – Hach Portable Spectrophotometer DR 2800 User Manual

Page 78

Advertising
background image

78

Advanced Operations

Select Conc. key
If an already defined formula, in this case C1, is to be taken into
account in the formula for C2, press Select conc.

Enter the number of the formula (e.g. 1 for C1) and press OK to
confirm.

Cn can now be linked with a mathematical operation.

Note: The Cn concentrations that are to be calculated are numbered in
sequence: C1, C2, C3, etc.

Note: When the first Cn formula has been defined, the Formula list is
automatically extended by Cn+1.

6.1.2.4 Enter variables

1. Highlight the Variables line in the data overview and press

Edit.

2. Select the variable to be edited, press Edit and use the

alphanumeric keypad to enter the data specified in the working
procedure (for F1, F2,

λ1, U1 etc.). Press OK to confirm each

entry.

Abbreviation of Variables:

F1: Factor 1

F2: Factor 2

λ1: Wavelength 1

U1: Conversion Factor 1 for the first chemical form

U2: Conversion Factor 2 for the further chemical form etc.

6.1.2.5 Save a free programming user program

1. Press Store to save the entered data. The data can be stored

under any data point (Measurement sequence, Formula, Timer,
etc.).

2. Press the PC & Printer icon to send the program data to a

printer or to a USB memory stick (connect the USB memory
stick to the USB interface first).

Note: The test data will be formatted in the preinstalled PrgData folder
as a .csv and as a .lst file.

3. To transfer the program data from one instrument to another

create a new folder on the USB memory stick named dbhlc
and copy the .lst files into this folder (see

section 6.7.2 on

page 98

).