Advanced operations – Hach Portable Spectrophotometer DR 2800 User Manual

Page 67

Advertising
background image

67

Advanced Operations

2. To enter the standard concentrations and corresponding

absorbance values in the displayed table, press the "+" symbol.
Use the alphanumeric keypad to enter the values.

Press OK and enter the corresponding absorbance value.
Press OK.

The entered data are displayed in the table. Repeat the
sequence for each data point to enter it.

3. To change a value in the table, highlight the appropriate line,

press the unit key (e.g. mg/L) or Abs and enter the changed
value via the alphanumeric keypad.

4. When the data have all been entered, press Graph.

Note: The correlation coefficient (r

2

) is shown on the left below the

axes.

5. The linear equation corresponds to the standard setting. Press

Next Curve to display the polynomial 2nd order curve. Press
Next Curve again, to display the polynomial 3rd order curve.

6. Press Force 0 to change the setting from Off to On. The curve

now passes through the origin of the coordinate system.

Note: This may have an adverse effect on the correlation coefficient
(r

2

).

7. Press Table to display the table again.

8. When the table has been completed and the curve type has

been chosen, press Done when the graph is displayed or Exit
when the table is displayed. Go to

section 6.1.1.4 on page 70

.