Section 1

Specifications

Specifications are subject to change without notice.

Performance Specifications

Operating Mode

Transmittance (%), Absorbance and Concentration

Source Lamp

Gas-filled Tungsten (visible)

Wavelength Range

340–900 nm

Wavelength Accuracy

± 1.5 nm

Wavelength Reproducibility

< 0.1 nm

Wavelength Resolution

1 nm

Wavelength Calibration

Automatic

Wavelength Selection

Automatic, based on method selection

Spectral Bandwidth

< 8 nm

Photometric measuring range

± 3.0 Ext in Wavelength Range 340–900 nm

Photometric Accuracy

5 mAbs at 0.0–0.5 Abs
1% at 0.50–2.0 Abs

Photometric Linearity

< 0.5%–2 Abs
< = 1% at > 2 Abs
with neutral glass at 546 nm

Stray Light

< 0.2% T @ 340 nm with KV450/3
< 0.1% T @ 340 nm with NaNO

2

Data storage

500 measured values (result, date, time, sample ID, user ID)

User programs

50

Physical and Environmental Specifications

Width

220 mm (8.6 in)

Height

135 mm (5.3 in)

Depth

330 mm (12.9 in)

Weight

4.06 kg (8.95 Ibs) without battery
4.38 kg (9.66 Ibs) with battery

Operating Requirements

10–40 °C (50–104 °F), max. 80% relative humidity (non-condensing)

Storage Requirements

-40–60 °C (-40–140 °F) max. 80% relative humidity (non-condensing)

Additional technical data

Mains connection

15 V- / 30VA
Plug-in power supply unit: (100–240 V/50–60 Hz)

Interfaces

Use only shielded cable with max. length of 3 m.
1 x USB type A
1 x USB type B

Enclosure Rating

IP 41 with closed lid
IP 42 with Protective Cover in place

Safety class

Safety class II