Operators Manual

2-4

Table 2-1. Description of Quantities

Quantity Description

Units

Comments

ADCF

Air Density

Correction

Factor

None

Calculated by instrument and depends on front panel
Temperature and Pressure settings.

QExposure

Charge during

exposure

C

Electrical charge delivered to readout in response to
radiation incident upon detector since the last
exposure.

QAccum

Accumulated

Charge

C

The sum of all charge accumulated due to exposures
since the RESET/MEASURE button was last pressed.

QAccum_previous

Previous

Accumulated

Charge

C

The sum of all accumulated charge not including the
current exposure since the RESET/MEASURE button
was last pressed.

QElectrometer

Electrometer

Charge

C

Factory calibrated electrical charge.

Q

Charge C

Calibrated

electrical

charge.

QScaleFactor

Charge Scale

Factor

C / C

Calibrates the charge measured by the electrometer.
Viewed on a Test Function screen.

QOffset

Charge

Leakage

C

The electrical charge measured when no incident
radiation is present.

I

Current A

Calibrated

electrical

current.

IAvg

Average

Current

A

The constant electrical current required to accumulate
charge Q over time TStart to T.

ICCF

Ion Chamber

Calibration

Factor

Ion

Chamber

Calibration

Units

/ A or C

Values are field programmable via Customization
Software.

ILeakage

Leakage

Current

A

The electrical current measured when no incident
radiation is present.

IScaleFactor

Current Scale

Factor

A / A

Calibrates the current. Viewed on a Test Function
screen.

RateMultiplier

Rate Units

Multiplier

s

Specifies the number of seconds the rate divisor is
equal to 1(s), 60(min), 3600(hr).

T

Q Sampling

Time

s

The time corresponding to Q and I.

Time

Effective Time

of Exposure

s

The time required to accumulate charge QExposure
with a constant electrical current IAvg.

TOffset

QOffset

Sampling Time

s

The time corresponding to QOffset and ILeakage.

QTimed

Timed Charge

C

The electrical charge delivered during the timer
interval.

QTimed_Predicted

Predicted

Timed Charge

C

The total amount of timed charge that should be
accumulated at the completion of the timer interval,
based on the observed delivery rate.