Appendix

Theory of Operation

B

B-11

B.5 Theory Of Operation

Basic Operating Equations

Quantity Description

Units Comments

Q

Timed

Timed Charge

C

The electrical charge delivered during the timer
interval.

Q

Timed_Predicted

Predicted
Timed Charge

C

The total amount of timed charge that should be
accumulated at the completion of the timer
interval, based on the observed delivery rate.

Leakage Measurements

Leakage measurements (Q

Offset,

I

Leakage

and T

Offset

) are only applicable when automatic leakage

compensation is enabled.

Timed Charge and Dose

This measurement accumulates charge only while the elapsed time is less than or equal to the timer
interval setting. The displayed value is held following the completion of the timer interval.

(

)

(

)

[s]

[s]

[A]

[C]

[C]

1

2

Leakage

1

2

Timed

T

T

I

Q

Q

Q

×

=

ADCF

ICCF

Q

Dose

Timed

Timed

Ч

Ч

=

[Units/C]

[C]

Q

1

represents the next sample of charge at time T

1

following the activation of the timer. Q

2

corresponds to

the interpolated charge at time T

2

, where T

2

is an integral number of seconds that is restricted to times

within the range: T

1

< T

2

≤ (T

1

+ Timer Interval Setting).

Predicted Timed Charge and Dose

This measurement extrapolates the timed charge (Q

Timed

) at the elapsed time (T

Elapsed

) that should be

accumulated at completion of the timer interval (i.e., TimerIntervalSetting) based on the rate of charge
delivery.

[s]

[s]

[C]

valSetting

TimerInter

T

Q

Q

Elapsed

Timed

icted

Timed_Pred

×

=

ADCF

ICCF

Q

Dose

icted

Timed_Pred

icted

Timed_Pred

Ч

Ч

=

[Units/C]

[C]

At the conclusion of the timer interval (i.e., T

Elapsed

equals the TimerIntervalSetting), the predicted timed

charge and dose are equal to the predicted timed charge and dose, respectively.