Troubleshooting and Service

Line Fuse Replacement

4


4-9

4.6 Line Fuse Replacement


Disconnect the Model 35040 Therapy Dosimeter
from other equipment along with the AC line power
and turn off power before continuing with this
section.

1. Position the instrument on its display face.



























Figure 4-1.

Model 35040 Therapy Dosimeter Fuse Replacement

2. Remove the fuse drawer. With your fingernails, squeeze together the two ends of the fuse drawer

and pull it out of its slot.

3. Replace the fuses with P/N 38476.

The use of different fuses may result in fire or
electrical shock in case of an internal fault.

4. Place the fuse drawer back into its slot. Press on the fuse drawer firmly until it snaps back into

place. Both sides of the fuse drawer should snap into place.

NOTE

WARNING