Gossen MetraWatt 702 User Manual

Page 31

Advertising
background image

GOSSEN METRAWATT GMBH

31

Test Sequence per VDE 0701, Part 240

Select connection, select test regulation VDE 0701, part 240, classify the DUT (SC I, II or III), combination testing X/– (yes/no)

Visual inspection

OK?

SC I: RPE

OK?

Short-circuit

at DUT?

Yes/No

No

Yes

Display results (for combination testing: additional display of differential resistance), save/print report

Yes

No

No

No

Test
OK?

Framed with dashed line:
The test is only run if it has been
activated in the initial program
window, or in the Setup menu under
Sequence ...

Switch line voltage to test socket, start function test

Start

function test

IL

OK?

Switch line voltage to test socket, start measurement

Reverse mains polarity, perform measurement again

Yes

Contact all exposed, conductive parts with the probe.
The test can be repeated as often as desired for various protective
conductor parts with the manual sequence*.

* If it is not clear whether or not all

exposed, conductive parts are
connected to one another or to the
protective conductor, testing can
be performed in the manual mode.

Switch DUT on