Gossen MetraWatt 702 User Manual

Page 42

Advertising
background image

42

GOSSEN METRAWATT GMBH

Power Supply

Line Voltage

207 V

… 253 V

Line Frequency

50/60 Hz

Power Consumption

Approx. 30 VA

for function test

Continuous max. 3600 VA,
power is conducted through the instrument only,
switching capacity

≤ 16 A

Electrical Safety

Safety Class

I per IEC 61010-1/EN 61010-1/VDE 0411-1

Nominal Voltage

230 V

Test Voltage

3.7 kV 50 Hz

Measuring Category

II

Fouling Factor

2

Safety Shutdown

Where differential current at the
DUT > 25 mA
Shutdown time: < 100 ms
Probe current: > 10 mA, < 1 ms

Electromagnetic Compatibility

Generic Standard

DIN EN 61326:2002

Mechanical Design

Display

Multiple dot matrix display, 128 x 128 pixels

Dimensions

LxWxH: 292 mm x 138 mm x 243 mm

Weight

Standard instrument: approx. 4.0 kg

Protection

Housing: IP 40
Terminals: IP 20

per

DIN VDE 0470 part 1 / EN 60529

Extract from table on the significance of IP codes

RS 232 Interface

Type

RS 232C, serial, per DIN 19241

Configuration

9600, N, 8, 1

Connection

9-pin subminiature socket connector

Interference Emission

Class

EN 55022

B

Interference Immunity

Test value

Feature

EN 61000-4-2

Contact/atmos. - 4 kV/8 kV A

EN 61000-4-3

10 V/m

C

EN 61000-4-4

Mains connection - 2 kV

B

EN 61000-4-5

Mains connection - 1 kV

A

EN 61000-4-6

Mains connection - 3 V

A

EN 61000-4-11

0.5 period / 100%

A

IP XY

(1

st

digit X)

Protection against

foreign object entry

IP XY

(2

nd

digit Y)

Protection against the

penetration of water

0

not protected

0

not protected

1

≥ 50.0 mm dia.

1

vertically falling drops

2

≥ 12.5 mm dia.

2

vertically falling drops with

enclosure tilted 15

°

3

≥ 2.5 mm dia.

3

spraying water

4

≥ 1.0 mm dia.

4

splashing water

Advertising