Gossen MetraWatt 702 User Manual

Page 41

Advertising
background image

GOSSEN METRAWATT GMBH

41

Multimeter Measurements for Devices with Feature F01

Reference Ranges

Line Voltage

230 V

±0.2%

Line Frequency

50 Hz

±0.1%

Waveshape

Sine (deviation between effective and rectified
value < 0.5%)

Ambient Temperature

+23

°C ±2 K

Relative Humidity

40

… 60%

Load Resistance

Linear

Nominal Ranges of Use

Line Voltage

207 V

… 253 V

Line Frequency

50 Hz

Line Voltage
Waveshape

Sine

Temperature

0

°C … + 50 °C

Ambient Conditions

Storage Temperature

– 20

°C ... + 60 °C

Operating Temperature – 10

°C ... + 50 °C

Accuracy Range

0

°C ... + 50 °C

Relative Humidity

Max. 75%, no condensation allowed

Elevation

Max. 2000 m

Deployment

Indoors, except within specified ambient
conditions

Influencing Quantities and Influence Error

Func-
tion

Measured Quantity Measuring Range /

Nominal Range of
Use

Reso-
lution

Open-
Circuit
Voltage
U

0

Short-
Circuit
Current
I

K

Internal
Resis-
tance
R

I

Measuring
Error

Intrinsic Error

Overload Capacity

Value

Duration

U

AC/DC

Voltage

0 ... 253.0 V

,

and

0.1 V

±

(2.5 % rdg. +5 d)

> 10 d

253 V

cont.

Low Voltage, SC III

±

(5% rdg. +10 d)

R

Resistance

0 ... 150.0 k

Ω

100

Ω

< 20 V –

1.1 mA

±

(1% rdg. +3 d)

253 V

cont.

I

Clip

Current with

Clip-On Current-

Voltage Transformer

WZ12C

0.000 ... 10.00 A

1 mA

1.5 M

Ω

±

(3% rdg. +10 d)

> 10 d
without clip-on device

253 V

cont.

0 ... 100 A

1 A

1.5 M

Ω

253 V

cont.

Temp.

Temperature

with Pt100/Pt1000

Sensor

– 200 ... – 50

°

C

1

°

C

< 20 V –

1.1 mA

±

(2% rdg. +1

°

C)

10 V

cont.

– 50.1 ... + 300.0

°

C

0.1

°

C

±

(1% rdg. +1

°

C)

10 V

cont.

+300 ... +850

°

C

1

°

C

±

(2% rdg. +1

°

C)

10 V

cont.

Influencing Quantity /
Sphere of Influence

Designation
per
DIN VDE 0404

Influence Error

± … % of Measured Value

Position Change

E1

Change in Test Setup Supply Power

E2

2.5

Temperature Fluctuation

E3

Specified influence error applies per
10 K change in temperature:

0

… 21

°

C and 25

… 40

°

C

1 in case of PE measurement

0.5 for all other measuring ranges

Current at Device Under Test

E4

2.5

Low-Frequency Magnetic Fields

E5

2.5

Impedance at Device Under Test

E6

2.5

Capacitance, Insulation Measurement

E7

2.5

Waveshape of Measured Current

E8

49

… 51 Hz

2 for capacitive load (for equivalent
leakage current)

45

… 100 Hz

1 (for contact current)

2.5 for all other measuring ranges

Advertising