11 APPENDIX B, EMC INFORMATION

GUIDANCE AND MANUFACTURER’S
DECLARATION

Electromagnetic Emissions—for all Equipment
and Systems

Guidance and manufacturer’s declaration—

electromagnetic emission

The Grafco Ultrasound Pocket Doppler is intended for use in the

electromagnetic environment specified below. Do not use the system in

environments which do not comply with the specifications listed below.

Emission

test

Compliance

Electromagnetic environment–

guidance

RF emission

CISPR 11

Group 1

The Grafco Ultrasound Pocket Doppler

uses RF energy only for its internal function.

Therefore, its RF emissions are very low

and are not likely to cause any interference

in nearby electronic equipment.

RF emission

CISPR 11

Class B

The Grafco Ultrasound Pocket Doppler

is suitable for use in all establishments,

including domestic establishments and

those directly connected to the public

low-voltage power supply network that

supplies buildings used for domestic

purposes.

Electromagnetic Immunity—for all Equipment
and Systems

Guidance and manufacture’s declaration—electromagnetic

immunity

The Grafco Ultrasound Pocket Doppler is intended for use in the

electromagnetic environment specified below. Do not use the system in

environments which do not comply with the specifications listed below.

Immunity

test

IEC 60601

test level

Compliance

level

Electromagnetic

environment—guidance

Electrostatic

discharge

(ESD) IEC

61000-4-2

±6 kV

contact ±8

kV air

±6 kV

contact ±8

kV air

Floors should be wood,

concrete or ceramic tile. If

floor is covered with synthetic

material, the relative humidity

should be at least 30% RH.