Graham Field GRAFCO 14423-8 User Manual

Page 7

Advertising
background image

14423-INS-LAB-RevD11

7

WARNING: Do not the use Pocket Doppler if the bat-

tery cover is open.

WARNING: Use only those probes provided by GF

Health Products, Inc. intended to be used with the

Grafco Pocket Doppler.

WARNING: Do not stretch the probe cable further

than 6.5 feet (2 meters).

WARNING: Electromagnetic Interference: Before op-

eration, ensure that the environment in which the de-

vice is operated is not subject to any source of strong

electromagnetic interference, such as radio transmit-

ters, mobile telephones, etc.

WARNING: The following safety checks/tests must be

performed at least once every two years or as speci-

fied in your facility's test and inspection protocol, by

a qualified person with adequate training, knowledge,

and practical experience:
• Inspect the equipment for mechanical and func-

tional damage.

• Ensure the safety labels are legible.

• Verify that the device functions properly as de-

scribed in this manual.

• Test according to the pregnant woman's leakage

current (IEC 60601-1/1988: Limit: 100 uA (B)).

The leakage current should never exceed the speci-

fied limit. Record the data and store with the Grafco

Pocket Doppler. If the device is not functioning prop-

erly or fails any of the above tests, contact GF Health

Products, Inc. Technical Support at the phone number

on the back cover, or your GF authorized distributor.