Electromagnetic Immunity—for all Equipment
and Systems that are not Life-Supporting

Guidance and manufacturer’s declaration—

electromagnetic immunity

The Grafco Ultrasound Pocket Doppler System is intended for use in the

electromagnetic environment specified below. Do not use the system in

environments which do not comply with the specifications listed below.

Immunity

test

IEC 60601

test level

Compliance

level

Electromagnetic

environment—guidance

Radiated

RF IEC

61000-4-3

3 V/m 80

MHz to 2.5

GHz

3 V/m

Portable and mobile RF

communications equipment

should be used no closer to any

part of the Grafco Ultrasound

Pocket Doppler, including

cables, than the recommended

separation distance calculated

from the equation applicable to

the frequency of the transmitter.

Recommended separation

distance

80 MHz to

800 MHz

800 MHz to

2.5 GHz

Where P is the maximum output

power rating of the transmitter

in watts (W) according to the

transmitter manufacturer and d

is the recommended separation

distance in meters (m).
Field strengths from fixed RF

transmitters, as determined

by an electromagnetic site

survey,

a

should be less than

the compliance level in each

frequency range.

b

Interference may occur in the

vicinity of equipment marked

with the following symbol: