During any test – MediSense PCx User Manual

Page 132

Advertising
background image

Questions? Refer to Page 10-13 For Customer Care Information.

10-7

During Any Test:

Symptom/Error
Message

Explanation

How To Respond

The barcode just scanned
was not read correctly.
Possible reasons are:

• The red scanner beam

did not pass across all
the bars of the code.

• The scanner was not

held between 3 and 12
inches from the barcode.

• The barcode was

damaged or poorly
printed.

• The barcode was not of

the type specified for
this ID or lot number.

1. Rescan the barcode or

manually enter the ID
or lot number.

2. Notify the system

administrator. The
problem may be the
printing of the barcode
or the setup.

The ID or lot number just
scanned or entered is too
long or too short,
according to format
defined during the setup
of this monitor.

1. Verify and re-enter the

ID or lot number using
the barcode scanner.

2. Manually enter the ID

or lot number.

3. If the problem persists,

notify the system
administrator.

The ID or lot number just
scanned or entered* is not
on the list of acceptable
IDs or lot numbers
defined for this monitor.

1. Verify and re-enter the

ID or lot number using
the barcode scanner.

2. Manually enter* the ID

or lot number.

3. If the problem persists,

notify the system
administrator.

The reason may also
be: Too Long

* Manual entry of strip lots with alpha-numeric barcodes is not supported.