What you see on the display what you do comments – MediSense PCx User Manual

Page 25

Advertising
background image

Questions? Refer to Page 10-13 For Customer Care Information.

4-3

7. With the contact bars

facing up, insert the test
strip into the test strip
port until it stops.

The monitor will prompt
the operator to apply the
desired control solution
to the test strip.

What You See
on the Display

What You Do

Comments

5. Scan or manually

enter* the test strip lot
number via the keypad,
then press Enter.

After the test strip lot
number is entered, the
monitor will prompt for
a test strip to be inserted
into the test strip port.

6. Open the foil test strip

packet at the notch and
tear up or down to
remove the test strip.

* Manual entry of strip lots with alpha-numeric barcodes is not supported.

Precision PCx Plus Test Strip
with TrueMeasure Technology
For Use with Precision PCx
v2.2 Only

Precision PCx Test Strip
For Use with Precision PCx
v2.0, v2.1 and v2.2.