Rs-232 only – Metrologic Instruments MS7120 User Manual

Page 30

Advertising
background image

26

T

ROUBLESHOOTING

G

UIDE

(

CONTINUED

)

RS-232 Only

SYMPTOMS

POSSIBLE CAUSE(S)

SOLUTION

Power-up OK
and scans OK
but does not
communicate
properly to the
host

Com port at the host is
not working or
configured properly

Check to make sure that the
baud rate and parity of the
scanner and the communication
port match and the program is
looking for “RS-232" data

Power-up OK
and scans OK
but does not
communicate
properly to the
host

Cable not connected
to the proper com port

Check to make sure that the
baud rate and parity of the
scanner and the communication
port match and the program is
looking for “RS-232" data

Power-up OK
and scans OK
but does not
communicate
properly to the
host

Com port not
operating properly

Check to make sure that the
baud rate and parity of the
scanner and the communication
port match and the program is
looking for “RS-232" data

The host is
receiving data
but the data
does not look
correct

The scanner and host
may not be configured
for the same interface
font

Check that the scanner and the
host are configured for the same
interface font

Characters are
being dropped

Intercharacter delay
needs to be added to
the transmitted output

Add some intercharacter delay
to the transmitted output by
using the MetroSelect

®

Programming Guide

MLPN

00-

02407 or the MetroSet2 program