Programming special functions, Setting the buzzer – Miele G 7883 User Manual

Page 41

Advertising
background image

3. Setting the buzzer

A buzzer can be set to signal the
following:

– End of program = continuous buzzer

– Error = rhythmic buzz

– Timer to allow sampling of the last

final rinse water
= intervals of 3 short buzzes and
2 seconds break

^ Turn the program selector to "STOP"

f.

^ Turn the machine off.
^ Press and hold "Display" ß and

"Start"

6.

At the same time,

turn the

lab washer on with the "On/Off"
button.
The displays shows the current
program status "P...".
The "Fill/Drain" indicator lights.

^ Press "Drying" 0 4 times. "E04"

appears in the display.

^ Turn the program selector to

"Inorganica". A number between "20"
and "27" appears in the display.
"20" =

Buzzer is turned off

"21" =

End of program
(factory setting)

"22" =

Error

"23" =

Error and End of program

"24" =

Test sample extraction

"25" =

End of program and
Test sample extraction

"26" =

Error and Test sample
extraction

"27" =

All functions
(End of program, Error,
Test sample extraction).

^ Press and hold "Display" ß until the

desired value appears in the display.

^ Press the "Start" 6 button. "SP"

appears in the display.

^ Press the "Start" 6 button again. The

change has been saved.

^ Please note the changed setting:

To turn the buzzer off:

^ Turn the program selector to "STOP"

f or press the "Display" ß button.

Programming special functions

41