Vacuum adjustment knob, Figure 1 medical suction unit, Using the suction unit – Invacare 636959 User Manual

Page 4: Caution, Connect the unit to power, Turn the power switch on (i), Clockwise to increase vacuum, Counter-clockwise to decrease vacuum, Using the suction unit caution

Advertising
background image

Aspirator Suction Unit

4

Part No 636959

FIGURE 1 Medical Suction Unit

Using the Suction Unit

CAUTION

Should fluid be aspirated back into the unit, send your suction unit to an

authorized service center for a detailed inspection as the vacuum pump may

have been damaged.

NOTE: For this procedure, refer to FIGURE 1 and FIGURE 2 on page 5.

NOTE: If the unit sustains a severe drop, validity of the gauge must be checked. Inspect if 
the float shut‐off is still functioning properly when the liquid in the collection bottle reaches 
the safe full level.

NOTE: Always transport unit with vacuum adjust knob rotated fully clockwise in case 
unit is dropped.

1. Ensure the suction tubing is connected to the labeled patient connector(outlet) 

on the collection bottle.

2. Ensure that all connections are secure and without leaks.

3. Connect the unit to power.

4. Turn the power switch On (I).

5. Adjust the vacuum level from 0 to 560 mmHg by turning the vacuum adjust 

knob: 

• Clockwise to increase vacuum. 

• Counter‐clockwise to decrease vacuum.

NOTE: Refer to the vacuum gauge while setting the desired level of vacuum. To accurately 
read the gauge, block the patient end of the hose or cap off the collection bottle and allow the 
gauge to reach a stable vacuum reading.

Vacuum

Adjustment

Knob

Vacuum Gauge

Collection

Bottle

Holder

Hydrophobic Filter

Connection

Tubing

Power Cord

Fuse

Elbow Connector

Lid

On (I)

Position

Off (O)
Position

Power

DETAIL “A”

Connection

Tubing