Troubleshooting, Problem, Solution – Invacare 636959 User Manual

Page 8: Suction unit does not turn on, Check the power cord for damage, Contact your invacare representative, Motor runs but no vacuum, Verify tubing connection security, Check for leaks or tubing kinks, Ensure that float shut-off is not activated

Advertising
background image

Aspirator Suction Unit

8

Part No 636959

Troubleshooting

NOTE: The suction unit contains no user‐serviceable parts except the fuse. 

Replacement Items/Accessories

PROBLEM

SOLUTION

Suction unit does not turn On. 1. Check that the plug is firmly fitted to the wall socket.

2. Ensure the wall outlet is live. Try plugging in another device such as
a lamp.
3. Check the power cord for damage.
4. Contact your Invacare representative.

Motor runs but no vacuum

1. Verify tubing connection security.
2. Check for leaks or tubing kinks.
3. Ensure that float shut-off is not activated.
4. Check for bottle leaks and cracks.
5. If these solutions do not work, contact your medical provider.

Low vacuum

1. Use vacuum adjust knob to increase vacuum level.
2. Check system for leaks.
3. Adjust vacuum adjust knob and release.

Overload protector shuts off
the motor frequently

Check number of motor operated appliances on same circuit and
adjust for aspirator unit.

Motor overheats and overload
protector shuts off motor

1. Turn power switch Off (O) and allow the motor to cool approxi-
mately 5 minutes. Restart compressor.
2. Inspect fuses and replace if necessary.
3. Call your medical provider if unit still does not start.

Blown fuse

Replace the fuse with the same type of fuse (Miniature Glass Fuse
5x20 mm, fast blow (quick acting)

F2A/250V).

PART NO.

DESCRIPTION

IRC1140

Collection jar with 6 ft. suction tubing

1148804

Tube and filter kit