Aspirator Suction Unit

Replacing the Hydrophobic Filter

NOTE: For this procedure, refer to FIGURE 3.

NOTE: Replace the hydrophobic filter when either of the following conditions occur: (1) the 
filter is dirty and the aspirator has lost suction, or (2) there is liquid in the hydrophobic 
filter assembly due to extended use or overflow.

NOTE: The hydrophobic filter should be replaced every two months. However, if overflow 
occurs, change the filter immediately. 

CAUTION

Use only Invacare recommended hydrophobic filters. DO NOT replace with

other types of bacteria filters.

1. Remove the elbow connector end of the hydrophobic filter assembly from the 

top of the collection jar.

2. Separate the opposite end of the hydrophobic filter assembly from the vacuum 

gauge.

3. Disassemble the hydrophobic filter assembly by disconnecting the two 

connection tubes from the hydrophobic filter housing.

4. Discard the filter.

5. Re‐assemble the hydrophobic filter assembly with a new filter. 

6. Re‐attach the hydrophobic filter assembly onto the collection bottle lid and 

vacuum gauge.

7. If necessary, twist the elbow connector to align with the collection jar.

FIGURE 3 Hydrophobic Filter Replacement

Hydrophobic

Filter

Collection Bottle Lid

Connection

Tubing

Connection

Tubing

Vacuum

Adjustment

Knob