Table 5 AC mA Test:

Step

Source

Frequency

Reading

1 58.00mA

50Hz 56.93

to

59.07

2 58.00mA

500Hz 56.93

to

59.07

3 580.0mA

50Hz 569.3

to

590.7

4 580.0mA

500Hz 569.3

to

590.7


6. Turn the rotary switch to mA .
7. Apply the inputs for steps 1-2 in Table 6.
8. For each input, compare the reading on the meter display to the reading in Table 6.
9. If the display reading falls outside of the range shown in Table 6, the meter does not meet specification.

Table 6 DC Current Test:

Step

Source

Reading

1 58.00mA

57.40

to

58.60

2 580.0mA

574.0

to

586.0


Testing the microamp Function
1. Turn the rotary switch to µA .
2. Apply the inputs for steps 1-2 in Table 7.
3. For each input, compare the reading on the meter display to the reading for your meter in Table 7.
4. If the display reading falls outside of the range shown in Table 7, the meter does not meet specification.

Table 7 DC µA Test:

Step

Source

Reading

1 580µA

574.0

to

586.0

2

5800µA

5740 to 5860


Testing the Frequency Function
To verify the accuracy of the meter’s frequency function, do the following:
1. Connect the calibrator to the VΩ and COM inputs on the meter.
Note: The accuracy of the calibrator’s frequency function must be appropriate for the specified
accuracy of the meter.
2. Set the rotary switch to Hz.
3. Set the function generator for the square wave voltage and frequency for steps 1-5 of Table 8.
4. Compare the reading on the meter display with the display reading shown in Table 8.
5. If the display reading falls outside of the range shown in Table 8, the meter does not meet specification.