IDEAL INDUSTRIES 61-320 User Manual

Page 7

Advertising
background image

Page 5


DC Current

Range

Resolution

Accuracy

Over voltage

protection

600.0µA 1mV

6000µA 1mV

60.00mA 10mV

600.0mA 100mV

±(1.0%+2 dgt)

600V rms1A (600) Fast
blow fuse


AC Current

Range

Resolution

Accuracy

Over voltage

protection

600.0µA 1mV

6000µA 1mV

60.00mA 10mV

600.0mA 100mV

±(1.5%+2 dgt)

50Hz ~ 500Hz *

600V rms

1A (600) Fast blow

fuse

Voltage Burden: µA: <4mV / µA
mA: 2V max.

* AC Conversion Type: Conversion type and additional specification are same as DC/AC Voltage.

Resistance

Range

Accuracy

Overload

protection

600.0Ω

6.000K Ω

60.00K Ω

600.0K Ω

±(0.7%+2 dgt)

6.000M Ω

±(1.0%+2 dgt)

60.00M Ω *

±(1.5%+2 dgt)

600V rms

Open circuit Voltage: -1.3V approx.
* < 2% of reading rolling.

Diode Check and Continuity

Range

Resolution

Accuracy

Overload

protection

10 mV

±(1.5% + 5 dgt)* 600V

rms

* For 0.4V ~ 0.8V
Max.Test Current: 1.5mA
Max. Open Circuit Voltage: 3V

Continuity: Built-in buzzer sounds when resistance is less than approximately 100 Ω. Response time is
approximately 100 msec.



Form number TM61320-2-4

Rev 2 September 2002

Advertising
This manual is related to the following products: