Actuator

Cantilever Arm

Spring

FIGURE 4 - REPOSITIONING/REPLACING THE

LOCKING MECHANISM IN THE

CANTILEVER ARM

Angled Portion of

Locking Mechanism

Locking

Mechanism

ARMREST PAD ADJUSTMENT/
REPLACEMENT (FIGURE 5)

Armrest Pad Adjustment

1. Remove the phillips screw from the rear of the

armrest pad.

2. Reposition the cantilever slide tube to one (1) of

five (5) positions.

3. Align the mounting hole in the cantilever slide tube

with the desired mounting hole from STEP 2.

4. Install the phillips screw UP through the desired

mounting hole in the armrest tube and the cantile-
ver slide tube mounting hole to secure the arm-
rest pad to the armrest tube.

5. Repeat STEPS 1-4 for opposite armrest pad if

necessary.

Armrest Pad Replacement

1. Remove the phillips screw from the rear of the

armrest pad.

2. Remove the phillips screw from the front of the

armrest pad.

3. Remove the EXISTING armrest pad from the arm-

rest tube.

4. Position the NEW armrest pad onto the armrest

tube and cantilever slide tube.

5. Secure the NEW armrest pad to the armrest tube

and cantilever slide tube with the two (2) phillips
screws.

REAR Phillips Screw

Cantilever Slide Tube

FRONT Phillips Screw

FIGURE 5 - ARMREST PAD ADJUSTMENT/

REPLACEMENT

Armrest

Tube

PROCEDURE 5

ARMS

A
R

M

S

REPLACING/REPOSITIONING THE
CANTILEVER ARM LOCKING
MECHANISM (FIGURE 4)

CAUTION

The locking mechanism is spring loaded. Place
your free hand over the end of the cantilever
arm to prevent the locking mechanism from
springing out of the arm.

1. Flip the cantilever arm up and out of the way.

2. Unscrew the actuator from the locking mechanism.

3. Remove the locking mechanism and spring from

the cantilever arm.

NOTE: Inspect the spring and replace if necessary.

4. Slide the NEW spring and locking mechanism into

the cantilever arm.

5. Position the angled portion of the locking mecha-

nism in one (1) of two (2) ways:

A. ANGLED PORTION FACING UP - Arm will

flip UP.

B. ANGLED PORTION FACING DOWN - Arm

will flip DOWN.

6. Push and hold the locking mechanism into the

cantilever arm.

7. Screw the actuator into the locking mechanism

mounting hole.

8. Flip the cantilever arm up/down to lock in place.

9. Pull down/up on the cantilever arm to ensure it

has locked in place.