Calibration Utility for MACINTOSH

2–13

Part No. 1H9659 September 1997

NOTE: If you want to print an updated test print, you can click on the

<<Back button to return to the Easy Cal Step 4 dialog box.

If cluster number 225 on the updated test print is the most neutral
gray, then the calibration is successful. Type 225 in the

Cluster

Number text box and go to step 3.

3. Click on the

Done button.

The printer calibration begins. When the calibration is complete, the
following dialog box appears.

4. Do one of the following:

D

If you do not want to make another test print and you do not want
to save the calibration table, click on the

Quit button to quit the

Calibration Utility.

or

D

If you want to save the calibration table, click on the

Proceed

button to access a document window. Refer to “Using the
Calibration Document Window” on page 2–24.

Saving Your Calibration Table

You may want to save the calibration table you just created to use as a
starting point for later calibrations.

D

Pull down the File menu and select

Save to save changes to the

current calibration table.

D

Pull down the File menu and select

Save As to access a selection box

in which you can specify a new name and location for the current
calibration table.