7 External control

7.1

External input/output control

■External control connectors (12 pin × 2)
MINI COMBICON Plug 12P

(PHOENIX CONTACT: MC 1.5/12-ST-3.5)
(Osada: OS-85-12P)

Compatible wiring (stranded)


Tightening torque: 0.22 to 0.25 Nm

Precautions for wiring

・ Carefully strip the cover so as not to damage the core wires.
・ Connect the core wires without twisting them.
・ Connect the core wires without soldering them; otherwise, they may break due to vibration.
・ After connection, do not apply stress to the cable.
・ Because of the terminal structure, if the wire is tightened by a counterclockwise rotation, the

connection will fail. In such cases, pull out the wire, check the terminal hole, and then connect it
again.

■Input/output terminal table


INPUT

Pin No

Signal name

Description

1

START 1

CH1 irradiation start signal

2

START 2

CH2 irradiation start signal

3

START 3

CH3 irradiation start signal

4

START 4

CH4 irradiation start signal

5

TYPE Chg1

Product type switching signal

6

TYPE Chg2

Product type switching signal

7

TYPE Chg3

Product type switching signal

8

UV CHECK

UV intensity control mode signal

UJ35 only

9

STOP

Irradiation stop signal, Error reset

10

START ALL

CH1-CH4 irradiation start signal

11

INTERLOCK

Interlock (normally ON)

12

COM

Common terminal for input/output signals

OUTPUT

Pin No

Signal name

Description

13

READY ALL

ON when irradiation is ready to start

14

READY1/BUSY1

ON when CH1 irradiation is ready to start/in progress

15

READY2/BUSY2

ON when CH2 irradiation is ready to start/in progress

16

READY3/BUSY3

ON when CH3 irradiation is ready to start/in progress

17

READY4/BUSY4

ON when CH4 irradiation is ready to start/in progress

READY/BUSY
singal is changed
over by changing the
initial setting mode.

18

ERROR

Error signal

19

ALARM

Warning signal (temperature/time warning)

20

COM

Common terminal for input/output signals

21

COM

Common terminal for input/output signals

22

FG

Frame gland

23

+5V

5 V DC output (for display or output signals)

24

COM

Common terminal for input/output signals

* Pin Nos. 12, 20, 21 and 24 are connected internally.

The ON time of input signal pulses must be 100 ms or more.

Size

Conductor section area

AWG#24-16

0.2 - 1.25 mm

2

6-7 mm

Wire stripping length