19

Compass calibration

If the Module was exposed to a strong magnetic field or was not
in use for a long time, the compass accuracy may diminish.
To bring the compass into working condition:

ƒ

Select

Cd 17

in the Setup mode

ƒ

Keeping the unit horizontally, perform one full rotation

around vertical axis within half a minute and then one full
rotation around vertical axis with device upside down
during half a minute.

ƒ

If one rotation does not restore calibration – perform two

rotations.

The normal compass functioning will be restored.
Depending on the version of the on-board software, during
calibration LCD will show either time countdown in seconds
from 59 to 1 or flashing word CAL.
To ensure azimuth measurement accuracy calibrate compass
after changing the battery.

Interface format

The unit can output measurement data via RS-232 port in either a
proprietary protocol or in the format accepted by PLGF/DALGR.
The selected format will be indicated on the LCD with display
names

PLGr

or

-PC-

.

To select the required data transfer format:

ƒ

Select

PLGr

(

-PC-

) in the Setup mode