NI 5422 Specifications

24

ni.com

Start Trigger

DDC CLK IN

Clocking
Format

DDC Data Output signals change on the rising edge of
DDC CLK IN.

Frequency
Range

10 Hz to 200 MHz

Input Duty
Cycle
Tolerance

40% to 60%

Table 8.

Specification

Value

Comments

Sources

1. PFI<0..1> (SMB front panel connectors)

2. PFI<2..3> (DIGITAL DATA & CONTROL front panel

connector)

3. PXI_Trig<0..7> (backplane connector)

4. PXI Star trigger (backplane connector)

5. Software (use function call)

6. Immediate (does not wait for a trigger). Default.

Modes

1. Single

2. Continuous

3. Stepped

4. Burst

Edge Detection

Rising

Minimum
Pulse Width

25 ns. Refer to t

s1

at NI Signal Generators Help»Devices»

NI 5422»NI PXI-5422»Triggering»Trigger Timing.

Table 7. (Continued)

Specification

Value

Comments