NI High-Speed Digitizers Getting Started Guide

Note

Windows Vista does not support some Traditional NI-DAQ (Legacy) devices. Refer to the

NI-SCOPE Readme to determine which operating system is compatible with your digitizer.

4.

Record the device number or device name assigned by MAX. You need this number when you
program your device.

(Traditional NI-DAQ (Legacy) Devices)

Select the device to see its properties in the configuration

view. The device number appears in the Value column. By default, the resource name for the
device is

DAQ::

n

, where

n

is the device number MAX assigned to your device. NI-SCOPE

devices that are configurable under Traditional NI-DAQ are not supported by the LabVIEW
Real-Time Module.

(NI-DAQmx Devices)

The assigned device name is appended to the device in its configuration

tree label. For example, after you install the digitizer, the device configuration tree label may
appear as

NI PXIe-5122:

"

dev1

", where

dev1

is the device name. When you develop

your application, the resource name is the device name assigned by MAX. Only the
NI-SCOPE devices that appear under NI-DAQmx Devices are supported by the
LabVIEW Real-Time Module.

Note

To avoid modifying existing applications that use a Traditional NI-DAQ (Legacy) device

number, change the assigned NI-DAQmx device name by right-clicking on the device, selecting
Rename, and entering the Traditional NI-DAQ (Legacy) device number used in your application.
For more information about device naming conventions, refer to the

niScope_init

function or the

niScope Initialize VI at NI High-Speed Digitizers Help»Programming»Reference.

5.

Perform a self-test on the device to verify installation.

(Traditional NI-DAQ (Legacy) Devices)

Right-click the device, select Properties, and click

Test Resources.

(NI-DAQmx)

Right-click the device and select Self-Test.

A dialog box indicates whether the device has passed the test.

Note

If the device does not pass the self-test, repeat the instructions in section

5. Installing the

Hardware

. If the device still does not pass, visit NI Technical Support at

ni.com/support

.

6.

Run the test panels on the device to verify the signal.

a.

Connect a signal to the digitizer and select the appropriate device parameters for this signal
such as range, input limits, sample rate, and sample mode.

b.

Access the test panel.

(Traditional NI-DAQ (Legacy) Devices)

Click Run Test Panels in the Properties window.

(NI-DAQmx)

Right-click the device and select Test Panels.

Note

For both types of devices, you can enable triggering by clicking the Advanced button on the

test panel.

All NI digitizers have self-calibration capabilities. You can access this feature for all NI digitizers
programmatically with NI-SCOPE and your ADE, or you can use the NI-SCOPE Soft Front Panel
(SFP). However, only NI-DAQmx devices can be self-calibrated using MAX.

7.

Exit MAX when you have finished configuring and testing the digitizer.