20

ni.com

Torque for screw terminals ............... 0.22 to 0.25 N · m

(1.95 to 2.21 lb · in.)

Ferrules ............................................. 0.25 mm

2

to 0.5 mm

2

Weight............................................... 50 g (1.8 oz)

Safety

Isolation Voltages

Connect only voltages that are within the following limits.

Channel-to-channel

Continuous ................................. 250 V

rms

,

Measurement Category II,
(Basic insulation)

Withstand.................................... 1,390 V

rms

, verified by a 5 s

dielectric withstand test

Channel-to-earth ground

Continuous ................................. 250 V

rms

,

Measurement Category II,
(Double insulation)

Withstand.................................... 2,300 V

rms

, verified by a 5 s

dielectric withstand test