NI 9265 Operating Instructions and Specifications
Full-scale output current
Minimum.................................... 20.1 mA
Typical ........................................ 20.6 mA
Maximum ................................... 21.0 mA
Output range ..................................... 0 to 20 mA
Compliance voltage .......................... 12 VDC max
Maximum load.................................. 600
Ω
Accuracy
Measurement Conditions
Percent of
Reading
(Gain Error)
Percent of
Range
*
(Offset Error)
Calibrated, typ (25 °C, ±5 °C)
0.11%
0.19%
Calibrated, max (–40 to 70 °C)
0.25%
0.4%
Uncalibrated, typ (25 °C, ±5 °C)
0.35%
1.4%
Uncalibrated, max (–40 to 70 °C)
0.85%
2.5%
*
Range equals 0 to 20.6 mA