Using the internal half-bridge completion, Temperature sensor connection, Table 3-1. maximum load per excitation channel – National Instruments SCXI -1122 User Manual

Page 31

Advertising
background image

Chapter 3

Signal Connections

© National Instruments Corporation

3-9

SCXI-1122 User Manual

Table 3-1. Maximum Load per Excitation Channel

Excitation Level

Maximum Load

3.333 V

225 mA

1 mA

5 k

Using the Internal Half-Bridge Completion

Your SCXI-1122 includes half-bridge completion for half-bridge and quarter-bridge setups. The
completion network consists of two 2.5 k

±

0.02% ratio tolerance resistors with a temperature

coefficient of

2 ppm/

°

C. These resistors are connected in series. To use the network, connect

the VEX/2 screw terminal on the terminal block to the negative input of the channel of interest.

CH+0

CH-0

VEX/2

VEX-

SCXI-1322 Terminal Block

120

Strain

Gauge

VEX+

120

Dummy

Resistor

Figure 3-7. Connecting a Quarter-Bridge Strain Gauge to Channel 0

Note: When using the half-bridge completion network with a quarter-bridge setup, you must

use an extra resistor to complete the bridge. Place this resistor on the terminal block
between the positive input channel and the negative excitation output.

Temperature Sensor Connection

Pins A1 and A3 are for connecting the isolated temperature sensor located on the SCXI-1322
terminal block for cold-junction compensation (CJC) of thermocouples connected to the
SCXI-1122. The connection is overvoltage-protected to 250 Vrms with power on and off.

Warning:

Exceeding the overvoltage protection on the temperature connections can damage
the SCXI-1122, the SCXIbus, and the DAQ board. National Instruments is

NOT

liable for any damages resulting from such signal connections.