The difference in ground potential between two instruments connected to
the same building power system is typically between 1 and 100 mV. This
difference can be much higher if power distribution circuits are improperly
connected. If a grounded signal source is improperly measured, this
difference might appear as a measurement error. The connection
instructions for grounded signal sources are designed to eliminate this
ground potential difference from the measured signal.
Input Modes
The following sections discuss single-ended and differential measurements
and considerations for measuring both floating and ground-referenced
signal sources.
Figure 2-5 summarizes the recommended input mode for both types of
signal sources.