External voltage, Counter – National Instruments NI USB-6008/6009 User Manual

Page 27

Advertising
background image

© National Instruments Corporation

27

NI USB-6008/6009 User Guide and Specifications

Compatibility ......................................... TTL, LVTTL, CMOS

Absolute maximum voltage range ......... –0.5 to 5.8 V with respect to GND

Pull-up resistor ....................................... 4.7 k

Ω to 5 V

Power-on state........................................ Input

Digital logic levels

External Voltage

+5 V output (200 mA maximum) .......... +5 V typical, +4.85 V minimum

+2.5 V output (1 mA maximum) ........... +2.5 V typical

+2.5 V accuracy ..................................... 0.25% max

Reference temperature drift ................... 50 ppm/°C max

Counter

Number of counters................................ 1

Resolution .............................................. 32 bits

Counter measurements ........................... Edge counting (falling-edge)

Counter direction.................................... Count up

Pull-up resistor ....................................... 4.7 k

Ω to 5 V

Maximum input frequency..................... 5 MHz

Minimum high pulse width .................... 100 ns

Level

Min

Max

Units

Input low voltage
Input high voltage
Input leakage current

–0.3

2.0

0.8
5.8

50

V
V

μA

Output low voltage (I = 8.5 mA)
Output high voltage

Active drive (push-pull), I = –8.5 mA
Open collector (open-drain), I = –0.6 mA, nominal
Open collector (open-drain), I = –8.5 mA, with external
pull-up resistor

2.0
2.0
2.0

0.8

3.5
5.0

V

V
V
V

Advertising