Hardware Overview

This chapter includes an overview of the NI 5911, explains the operation of
each functional unit making up your NI 5911, and describes the signal
connections. Figure 2-1 shows a block diagram of the NI 5911.

Figure 2-1. NI 5911 Block Diagram

Differential Programmable Gain Input Amplifier (PGIA)

The analog input of the NI 5911 is equipped with a differential
programmable gain input amplifier. The PGIA accurately interfaces to and
scales the signal presented to the analog-to-digital converter (ADC)
regardless of source impedance, source amplitude, DC biasing, or
common-mode noise voltages.

Protect/

Calibration

Mux

Calibration

Generator

PGIA

Noise

Shaper

Reference
Clock

Data

A/D Converter

100 MHz, 8-Bit

Digital Signal

Processor

Digital IO

Connector

Analog Input

Connector

AC/DC Coupling

Capture

Memory

Timing IO/

Memory Control